CPET Calibration Ranges

Modified on Tue, 22 Oct at 10:53 AM

 

 

Error

Warning

Good

Warning

Error

O2 gain factor

< 0.8

0.8 - 0.85

0.85 – 1.15

1.15 – 1.25

> 1.25

CO2 gain factor

< 0.8

0.8 – 0.9

0.9 – 1.1

1.1 – 1.2

> 1.2

 

 

 

 

 

 

O2 delay

 

 

< 500

500 – 600

> 600

CO2 delay

 

 

< 450

450 – 550

> 550

 

 

 

 

 

 

T90 O2

 

 

< 150

150 – 200

> 200

T99 O2

 

 

< 500

500 – 750

> 750

T99 O2 orig.

 

 

< 750

750 – 1000

> 1000

T90 CO2

 

 

< 150

150 – 200

> 200

T99 CO2

 

 

< 500

500 – 750

> 750

 

 

You can also check the pressure of the O2 cell and CO2 sensor.

 

Remove all tubes from the Ergostiks (front and back)

 

Gas Valve Open / Cal Gas Valve Closed

Range

CO2 Cell Pressure

20-40 below ambient pressure

O2 Cell Pressure

70-90 below ambient pressure

Gas Valve Closed / Cal gas Valve Closed

Range

CO2 Cell Pressure

50-70 below ambient pressure

O2 Cell Pressure

100-120 below ambient pressure

 

Set the pump speed to 0

CO2 Cell Pressure

ambient pressure +/- 10

O2 Cell Pressure

ambient pressure +/- 10

 