CPET Calibration Ranges

Modified on Tue, 16 Jun at 12:58 PM




You can also check the pressure of the O2 cell and CO2 sensor.

 

Remove all tubes from the Ergostiks (front and back)

 

Gas Valve Open / Cal Gas Valve Closed

Range

CO2 Cell Pressure

20-40 below ambient pressure

O2 Cell Pressure

70-90 below ambient pressure

Gas Valve Closed / Cal gas Valve Closed

Range

CO2 Cell Pressure

50-70 below ambient pressure

O2 Cell Pressure

100-120 below ambient pressure

 

Set the pump speed to 0

CO2 Cell Pressure

ambient pressure +/- 10

O2 Cell Pressure

ambient pressure +/- 10

 