Using a 4 Metre Tubeset in BlueCherry

Modified on Wed, 16 Mar, 2022 at 3:13 PM

With the 4 metre version you may have to correct the expected delay in BLUE CHERRY a bit to make the synchronization work.

 

This cannot be done in the settings of BLUE CHERRY, but we have a config file that you can import (run the bat file). 

By default the delay for the 4 meter tube is set to 1000 ms, but probably you have to set the value to 900 ms. So you can set the value for Value in the xml file to 900, save the file, close BLUE CHERRY and run the bat file. Then you can try the synchronization again.

 

The expected delay = delay during synchronization – delay during gas calibration


The expected delay is the value you have to enter in the xml file


The delay during synchronization is the value shown in the right value panel during the synchronization


The delay during gas calibration is shown in the table of the gas calibration (if necessary, you have to add this parameter to the table)



Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons

Feedback sent