With the 4 metre version you may have to correct the expected delay in BLUE CHERRY a bit to make the synchronization work.
This cannot be done in the settings of BLUE CHERRY, but we have a config file that you can import (run the bat file).
By default the delay for the 4 meter tube is set to 1000 ms, but probably you have to set the value to 900 ms. So you can set the value for Value in the xml file to 900, save the file, close BLUE CHERRY and run the bat file. Then you can try the synchronization again.
The expected delay = delay during synchronization – delay during gas calibration
The expected delay is the value you have to enter in the xml file
The delay during synchronization is the value shown in the right value panel during the synchronization
The delay during gas calibration is shown in the table of the gas calibration (if necessary, you have to add this parameter to the table)