How to Perform a Flow Calibration

Modified on Tue, 10 Dec, 2019 at 12:56 PM

1. Open Blue Cherry

2. Click on the ‘Calibration’ button on the left.



3. Make sure the Syringe Volume is set to 3 Litres, do this by clicking the ‘Syringe Volume’ button on the right.



4. Click ‘Record Volume calibration’ button on the right.



5. Follow instructions along the top of the calibration window. Blue Cherry will need 3 successful In & Ex strokes at 3 different flow rates Slow, Medium and Fast.



6. Once the calibration is completed click the ‘Save’ button in the top left.



7. The performed calibration will appear as the most recent result in the tread graph and at the top of the table below the graph, if the result has a green tick next to it in the table this indicates that the calibration has been successful. We recommend performing a Flow calibration before every test for optimised results.